Operation and use

Aspirated liquids and air are conveyed to the manifold (4) through
the tip and lifted hose and into the canister through tube 3A; inside
the canister liquids are separated from air.
The air is driven all the way to the motor (through pipe 3B) and
is then ex haust ed, while liquids are heavier and are collected at
the bottom of the canister.
Aspi-Jet 6 γ is equipped with a canister (5) whose capacity allows
8/10-hour continuous service before reaching max i mum level,
therefore the canister needs emptying every evening after work.
Aspi-Jet 7 γ has automatic drainage: a valve (6) located at the
bottom of the canister is kept close by depression when suction is
performed; on the contrary, when all terminals are on their seats
suction stops and liquids are drained out of the canister. In case
the canister should fi ll up during surgical operations, the probes
(7) will sense maximum level at about 3/4 of the canister (shortest-
probe level); the elec tri cal circuit opens and suction stops, while
a yellow-light indicator located on the front panel warns that the
canister is full. For Aspi-Jet 6 γ you need to switch the unit off and
empty the canister manually; for Aspi-Jet 7 γ , as mentioned above,
the draining valve opens and the draining pump (8) starts working.
In a few seconds the canister is empty and suction starts again
automatically. During surgical operations, foam build-up, caused
by blood and aspirated air, can reach probe level causing the unit
to stop; in this case we suggest the use of our solid anti-foaming
(directions for use are inside the package - Fig. B1).
In case some failure (clogging of cooling system, breakdown etc.)
should cause overheating of the motors - >120 °C for Uni-Jet
75 suction unit and >90 °C for draining pump (8) - a thermal
device rated at a fi xed temperature opens the circuit and resets
it automatically, when the temperature of the windings is back to
normal. Should this happen, identify and eliminate the cause.

Notice

Prior to starting any servicing operation on appliances
that have been used, clean with Puli-Jet plus a few times
as explained in the maintenance section. Use disposable
GLOVES (fi g. B3), GOGGLES, MASK and OVERALL.

Maintenance and cleaning

Electrical shock risk:
also 230 V can be lethal
Biological danger, risk of
infections from epidemic diseases
High temperature
Compulsory direction of fl ow and of rotation
16
Figura B3
9
4
Figura C
11
10
Figura D
A
B
Figura E