The sensors are connectable and can be removed for
maintenance work.
Do not touch or shock the sensors.
Humidity sensor
The dirt which gathers on the sensor must be water-
soluble. The sensor can be cleaned with
demineralised/destilled water or ISOPROPANOL (and
then rinsed in water). When cleaning the sensor the
following safety measures must be observed:
- The water temperature should not exceed +60 ° C.
- Pour the water into a spray bottle.
- Sprinkle the sensor lightly with water.
- Dry the sensor, preferably in nitrogen at a low
pressure. It can also be dried in air.
- Wait 24 hours before re-using or testing the sensor.
- Calibration should be effected after the sensor has
been cleaned. We, therefore, recommend the use of a
control and calibration set.
- Damaged sensors (e. g. through toxic gas or abraison)
can be replaced by new sensors.
Pt100/NTC
Clean in water or with a soft cloth.
Cleaning the sintered cap
The sintered filters are best cleaned with compressed
air, which is blown from the inside outwards.
Cleaning the sensor