3. Specifications
Tolerance values of the thermocouples per IEC 60584 / ASTM E230
Type
Tolerance value per Class
K
IEC 60584-1
N
ASTM E230
J
IEC 60584-1
ASTM E230
E
IEC 60584-1
ASTM E230
1) ItI is the value of the temperature in °C without consideration of the sign
2) The greater value applies
Limited error with selected temperatures in °C for thermocouples types K and N
Temperature
Tolerance value IEC 60584-1
(ITS 90)
Class 1
°C
°C
0
±1.5
100
±1.5
200
±1.5
±1.5
300
400
±1.6
500
±2.0
±2.4
600
700
±2.8
800
±3.2
±3.6
900
1,000
±4.0
1,100
-
1,200
-
Other thermocouple types on request
For further specifications see WIKA data sheet TE 65.59 or Technical information IN 00.23
and the order documentation.
WIKA operating instructions model TC59-V (V-PAD
Temperature range
1
-40 ... +1,000 °C
2
-40 ... +1,200 °C
Special
0 ... +1,260 °C
Standard
0 ... +1,260 °C
1
-40 ...
+750 °C
2
-40 ...
+750 °C
Special
0 ...
+760 °C
Standard
0 ...
+760 °C
1
-40 ...
+800 °C
2
-40 ...
+900 °C
Special
0 ...
+870 °C
Standard
0 ...
+870 °C
Class 2
°C
±2.50
±2.50
±2.50
±2.50
±3.00
±3.75
±4.50
±5.25
±6.00
±6.75
±7.50
±8.25
±9.00
)
®
Tolerance value
±1.5 °C or 0.0040 ∙ | t |
±2.5 °C or 0.0075 ∙ | t |
±1.1 °C or ±0.4 %
±2.2 °C or ±0.75 %
±1.5 °C or 0.0040 ∙ | t |
±2.5 °C or 0.0075 ∙ | t |
±1.1 °C or ±0.4 %
±2.2 °C or ±0.75 %
±1.5 °C or 0.0040 ∙ | t |
±2.5 °C or 0.0075 ∙ | t |
±1.0 °C or ±0.4 %
±1.7 °C or ±0.5 %
1) 2)
EN
9