3. Specifications / 4. Design and function

3. Specifications

Specifications
Nominal size
Accuracy class
Scale ranges
Max. working
pressure
(static pressure)
Overpressure safety
Ingress protection
Measuring chamber
with pressure
connection (wetted)
For further specifications see WIKA data sheet PM 07.14 and the order
documentation.

4. Design and function

Description
Pressures p
and p
1
by the magnetic piston under pressure(or magnetic piston and separating
diaphragm for model 700.02).
The difference in pressure causes an axial movement (deflection) of the piston
supported by the compression spring.
The deflection resulting from this is taken up by a ring magnet found on the
instrument pointer and appropriately displayed. This design combines the
advantages of complete separation of the measuring system and the indication
while eliminating any leakage to the exterior.
For model 700.01 the volume flow from the ⊕ measuring chamber to the ⊖
measuring chamber is minimised by the constructive design and will not interfere
with standard processes.
These instruments are mainly intended for gas/air supply and preparation appli-
cations where no magnetic particles are involved.
WIKA operating instructions differential pressure gauge with piston
Model 700.01
80
± 3 % of full scale value with
increasing differential pressure
0 ... 400 mbar to 0 ... 10 bar
optionally 100, 250 or 400 bar
either side to maximum working pressure
(exception for model 700.02, scale ranges 0 ... 160 mbar and
0 ... 250 mbar: Overpressure safety up to 50 bar)
lP 54 (EN 60529 / lEC 529)
Stainless steel 1.4571, 2 x G ¼ female, on the right and left side,
in-line (EN 837-1 /7.3)
are given in the ⊕ and ⊖ measuring chambers, separated
2
Model 700.02
± 5 % of full scale value with
increasing differential pressure
0 ... 160 mbar to 0 ... 2.5 bar
100 bar
(scale ranges 0 ... 160 mbar and
0 ... 250 mbar: 50 bar)