12. MESSAGES OF THE ALARM AND ERROR DISPLAYS

MESSAGE
CAUSE
The sterilization cycle was interrupted by the operator pressing the
ERROR100
Low pressure in the chamber while sterilizing
ERROR120
High pressure in the chamber while sterilizing
ERROR125
The pressure did not increase properly during the preheating phase or the machine could
not discharge the pressure at the end of the sterilizing cycle. If this is happening even if the
ERROR140
water is in the chamber, dry and empty the sterilization chamber, reset the alarm, let the
water in and restart the cycle up
The power was disconnected while the sterilization cycle
ERROR199
The cycle was started up without let water in the chamber for at least 20 sec. Dry and
ERROR400
empty the sterilization chamber, reset the alarm, let the water in and restart the cycle up
The cycle was started up but 5 min went from the water inlet in the chamber. Dry and
ERROR401
empty the sterilization chamber, reset the alarm, let the water in and restart the cycle up
The cycle was started up but the machine was hot. After the door
let the water out of the chamber.
ERROR450
Reset the alarm, wait until when the machine is cold, Dry and empty the sterilization
chamber, let the water in and restart the cycle up
During the preheating phase, in 40 minutes was not reached pressure value established for
sterilization. If this is happening even if the water is in the chamber, dry and empty the
ERROR700
sterilization chamber, reset the alarm, let the water in and restart the cycle up
Failure in the device for the pressure level
ERROR760
The pressure did not go down properly during the drying phase.
ERROR800
If an error message is shown the sterilization cycle has failed. It is necessary to repeat all the preparing and
sterilization phases.
If an error message appeared on the display during the sterilization phase, the cycle failed and the
instruments in the chamber were not sterilized.
To reset the error message push the
If the problem is coming out even if you carried out the listed operation, please contact
the MEDILINE assistance +39 0522 - 942997
MEDILINE IS NOT ACCEPTING ANY RESPONSIBILITY FOR TECHNICAL
INTERVENTIONS WHICH WERE CARRIED BY UNAUTHORIZED TECHNICIANS.