Valor 3000 Series Scales
Figure 2-3. Power Connection and Sealing
.

2.5 Initial Calibration

When the scale is first installed, it should be calibrated to ensure accurate weighing
results.
Press and hold the ENTER/Cal button until CAL is displayed. When the button is
again pressed and released, the display shows –C- while zero reading is stored.
Next, the display shows the calibration weight value in grams. Place the specified
calibration weight (sold separately) on the pan. See Table 2-1 for appropriate
weights. The display shows –C- while the reading is stored. The display returns to
the previous application mode and scale is ready for use.
TABLE 2-1. CALIBRATION WEIGHTS.
MODELS
V31XH202
V31XH2
V31X3
V31X6
V31XW301
V31XW3
V31XW6
Calibration
200g
2 kg
3 kg
6 kg
300 g
3 kg
6 kg
Weights

3. OPERATION

3.1 Weighing Mode

Use this mode to measure the weight of a sample in the selected unit of measure.
1. Press MODE/UNIT/Menu until the desired unit icon is displayed.
2. Put an empty container on the pan, then press O/T/On/Off to tare the container
weight.
3. Add material to the container. The display shows the weight of the material.