NOR disinfectant NOR alcohol gets inside the connector.
3
Dampen a clean cloth with a mild liquid soap solution. Wipe the outside
of the MiniLink.
4
Rinse the MiniLink under warm tap water but do NOT get water inside
the connector. If you get water inside the connector, shake the water out
and allow it to air dry.
5
Apply 3-4 drops of a quaternary ammonium compound disinfectant (for
example, Cavicide
6
Hold the MiniLink and rinse with 70% Isopropyl alcohol.
7
Disconnect the tester from the MiniLink.
8
Place the MiniLink on a clean, dry, non-shedding cloth and air dry.
9
Once it is completely dry, place it in a sealed bag labeled with the
cleaning date (see label example below).
Note: If there is any blood inside the connector, the MiniLink must be
discarded. Dispose of the MiniLink according to the local regulations
for battery disposal (non-incineration).
Storage
If you do not plan to use your MiniLink in the next 30 days, connect it to the
charger for storage.
®
) on a clean, dry cloth and wipe the MiniLink.
Example: Multiple-use cleaning label
Device:
Date:
Method of Decontamination:
Wash:
Disinfectant: