Answer: Check the connection. If you still do not see a green light
flash, fully recharge the MiniLink battery. Test the MiniLink with the
tester. If you still do not see a green light flash, call the HelpLine or
your local representative. It may be time to replace your MiniLink.
Cleaning the MiniLink
Cautions: The charger and the tester are NOT water-tight. Do
NOT immerse in water.
Do NOT discard the MiniLink in a medical waste
container or otherwise subject it to incineration.
MiniLink contains a battery which may explode upon
incineration.
For single-patient use
1
Wash your hands thoroughly.
2
Attach the tester to the MiniLink to make sure that NEITHER water
NOR anti-bacterial hand sanitizer gets inside the connector.
3
Dampen a clean cloth with mild liquid soap and warm water. Wipe
the outside of the MiniLink.
4
Rinse the MiniLink under warm tap water but do NOT get water
inside the connector. If you get water inside the connector, shake
the water out and allow it to air dry.
5
Using an antibacterial hand-sanitizer (available at a local
drugstore) on a clean, dry cloth, wipe the MiniLink's surface.
Do NOT get any hand-sanitizer inside the connector. Repeated
exposure to the hand-sanitizer could damage the connectors and
affect the MiniLink's performance as a result. If you get hand-
sanitizer inside the connector, allow it to air dry.
6
Disconnect the tester from the MiniLink.
7
Place the MiniLink on a clean, dry cloth and air dry for 2-3
minutes.
For multiple-patient use
1
Wash your hands thoroughly.
2
Attach the tester to the MiniLink to make sure that NEITHER water
- 9 -