8.2 diagnostic display indicators and Control

Service
1 - Cable or handpiece issue.
2 - Unit overheating. Stop use and
contact Cavitron Care.
Icon will appear when water filter needs to
be replaced. Once replaced, press and hold
to reset indicator.
Purge mode is used to flush water lines.
Runs automatically for 2 minutes.
This screen shows when purge
is activated.
• S cale Icon - press
to return to main
screen
• D epress foot pedal
to return to main
screen
• M ain screen will
appear when
countdown is
complete
Rinse mode is for use during an ultrasonic
scaling procedure when lavage is desired
with no cavitation.
This screen shows when rinse is activated.
Depress foot pedal to begin rinse mode
cycle.
• Scale Icon - press
to return to main
screen
Press & hold until settings
screen appears (see pg. 9).

Main Screen

8
Foot Pedal Battery
Indicates battery life of foot pedal.
Preset Power Modes-
Programmable
Press and hold until sound
indicator is heard.
Lock / Unlock Icons
Press and hold to lock or unlock
screen. Screen appears dimmed
when locked.
Power Level Icon
Adjust number to desired power
level by swiping in an upward or
downward motion.
• P ower levels set
between 05 and
35 are within Blue
Zone range