The fault is shown by an error code in the display (C) as following if the error occurs, e.g. Er 4.
Proceed as follows in such cases:
F Switch the device switch (A) off.
F Remove the stirrer tool and remove the instrument from the assembly.
F Reduce the speed and switch on (instrument switch (A)) the device without the stirrer tool.
Error code Cause
Er 2
Motor current sensor fault
Er 3
Temperature inside instrument
is too high
Er 4
Motor blockage or overload
Er 8
Speed sensor fault or overload motor off
Er 21
Safety relay fault
If the actions described fails to resolve the fault or another error code is displayed then take one of the
following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.
In accordance with IKA
®
warranty period is 24 months. For claims under
the warranty please contact your local dealer. You
may also send the machine direct to our factory,
enclosing the delivery invoice and giving reasons
for the claim. You will be liable for freight costs.

Error codes

Effect
motor off
motor off
motor off
motor off

Warranty

warranty conditions, the
Solution
- Switch off the instrument
- Switch off the instrument and allow it
to cool down
- Switch off the instrument
- Decrease the load on the motor and
restart again
- Switch off the instrument
- Switch off the instrument
The warranty does not cover worn out parts, nor
does it apply to faults resulting from improper use,
insufficient care or maintenance not carried out in
accordance with the instructions in this operating
manual.