Specifications in normal use (except during RECORD).
RANGE
WORKING
RANGE
SPECIFIED MEASURE-
MENT RANGE
MEASUREMENT ERROR
rdg : reading
Remark: this error is given (±).
The current detection threshold is 50 mA.
Stability of digital display (up to 4000 counts): 2 counts
Measurement resolution on digital display: 1 count
Display response time:
analog: from 100 to 150 ms to 95 %.
digital: 500 ms to 90 % / 2 s to 99 %.
Specifications on "RECORD 100 ms" mode (MAX, MIN, AVG and present value)
The specifications of the present value and the average value (AVG) are identical to those
in normal use.
Specifications of MAX and MIN values:
Accuracy: identical to normal mode.
Stability, repeatability: 10 counts.
MAX/MIN measurement (for a signal in the measurement range):
200 ms to 300 ms for currents above 8 A peak approx.
1 to 2 s for currents under 8 A peak (MAX and MIN taken on the current value).
Specifications in PEAK and RECORD PEAK
Resolution of digital display:
20 counts up to 4000 counts.
40 counts from 4000 to 10 000 counts.
RANGE
WORKING
RANGE
SPECIFIED MEASURE-
MENT RANGE
MEASUREMENT ERROR
rdg : reading
40.00 A
50 mA
100 A
300 mA
99.99 A
2 % rdg + 20 pts
40.00 A
300 mA
100 A
400 mA
99.99 A
3 % rdg + 30 pts
400.0 A
100 mA
700 A
100 A
400 A
2 % rdg
400.0 A
300 mA
1000 A
100 A
600 A
3 % rdg
50
400 A
700 A
5 % rdg
600 A
1000 A
8 % rdg

Werbung