Ac/dc-amperemeter
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

4.2. AC Current Measurements
Note:
Always start with the highest measuring range.
1.) Choose with the rotary switch the 5mA / 50mA / 500 mA resp.
10A position.
2.) Insert the black test lead into the COM socket and for current
measurements < 500mA the mA-socket. For current
measurements > 500mA use the 10A input socket.
3.) Switch OFF or disconnect the circuit to be measured from all
power sources and connect the multimeter in series with the
conductor in which the current to be measured flows.
4.) Switch ON the circuit and read the measured value from the
analogue display.
WARNING!
*
This unit is equipped with a 0.5 amp fuse for ranges < 500mA
and a 10 amp fuse for the 10 A range.
*
To prevent a risk of damage to the device and/or serious injury,
note maximum input values.