8. Maintenance, cleaning and recalibration
8.1.1 Battery replacement
DANGER!
Danger to life from explosion
Through working in flammable atmospheres, there is a risk of explosion
which can cause death.
▶
Only use the batteries listed, see chapter 3.7.3 "Permitted batteries"!
▶
Do not use rechargeable batteries!
▶
The instruments must not be opened in hazardous areas!
▶
Only replace the batteries outside the hazardous area
▶
Always replace all three batteries together!
▶
The battery cover must be closed and locked in place!
▶
Secure the battery cover with the three screws!
▶
Ensure the correct polarity.
Procedure
1. Turn off the instrument and lay it face down.
2. Loosen the three screws of the battery compartment, see Fig. 1 "Position of the
battery compartment".
3. Remove the battery cover.
4. Insert the three AA-sized batteries in correct polarity. Only use permitted batteries,
see chapter 3.7.3 "Permitted batteries".
5. Put the battery cover on and screw it tight with the three screws.
⇒
Tighten the screw on the top first.
Fig. 1 - Position of the battery compartment
If the instrument is not used for a long time, remove the batteries.
WIKA operating instructions precision digital pressure gauge CPG1500
EN