8. Maintenance, cleaning and recalibration

8.1.1 Battery replacement

DANGER!
Danger to life from explosion
Through working in flammable atmospheres, there is a risk of explosion
which can cause death.
Only use the batteries listed, see chapter 3.7.3 "Permitted batteries"!
Do not use rechargeable batteries!
The instruments must not be opened in hazardous areas!
Only replace the batteries outside the hazardous area
Always replace all three batteries together!
The battery cover must be closed and locked in place!
Secure the battery cover with the three screws!
Ensure the correct polarity.
Procedure
1. Turn off the instrument and lay it face down.
2. Loosen the three screws of the battery compartment, see Fig. 1 "Position of the
battery compartment".
3. Remove the battery cover.
4. Insert the three AA-sized batteries in correct polarity. Only use permitted batteries,
see chapter 3.7.3 "Permitted batteries".
5. Put the battery cover on and screw it tight with the three screws.
Tighten the screw on the top first.
Fig. 1 - Position of the battery compartment
If the instrument is not used for a long time, remove the batteries.