Ex Marking; Battery Replacement; Standards Applied For Certification - WIKA CPG1500 Betriebsanleitung

Vorschau ausblenden Andere Handbücher für CPG1500:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

3. Safety
Symbols
Before mounting and commissioning the instrument, ensure you read
the operating instructions!
This marking on the instruments indicates that they must not be disposed of
in domestic waste. The disposal is carried out by return to the manufacturer
or by the corresponding municipal authorities (see EU directive 2012/19/EU).

3.7 Ex marking

DANGER!
Danger to life due to loss of explosion protection
Non-observance of these instructions and their contents may result in the
loss of explosion protection.
Observe the safety instructions in this chapter and further explosion
instructions in these operating instructions.
Observe the information given in the applicable type examination
certificate and the relevant country-specific regulations for installation and
use in hazardous areas (e.g. EN/IEC 60079-14).
The precision digital pressure gauge is not intended for use with
flammable substances. It is only suitable for installation in locations
providing adequate protection against intrusion of solid foreign objects or
water which can impair safety.
Only use the batteries listed, see chapter 3.7.3 "Permitted batteries"!
Only replace the batteries outside the hazardous area, see chapter

8.1.1 "Battery replacement"!

Temperature range: -10 ... +50 °C

3.7.1 Standards applied for certification

IECEx BVS 16.0033X
IEC 60079-0:2011 (6th edition) + Corrigendum 1:2013
IEC 60079-11:2011 (6th edition)
IEC 60079-26:2014 (3rd edition)
WIKA operating instructions precision digital pressure gauge CPG1500
BVS 16 ATEX E043 X
EN 60079-0:2012 +A11:2013
EN 60079-11:2012
EN 60079-26:2015
EN
11