Emergency release

During a power failure the lid cannot be unlocked by motor. An emergency release has to be executed by hand.
For emergency release disconnect the centrifuge from the mains. In the case of permanently connected
devices, switch off the mains switch to disconnect the mains supply of the device in the building installation
and secure it against being switched on again, e.g. by locking the switch.
Open the lid only during rotor standstill.
See figure on page 2.

Switch off the switch on the front panel (switch position "0").

Look through the window in the lid to be sure that the rotor has come to a standstill.

Insert the hexagon socket head wrench horizontally into the drilling (Fig. 1, A) and carefully rotate by half a turn
in clockwise direction until the lid can be opened.

Pull the hexagon socket head wrench out of the drilling again.

After turning the centrifuge on again, press the button
the normal position (opened).
35

Maintenance and servicing

The device can be contaminated.
Pull the mains plug before cleaning. In the case of permanently connected devices, switch off the mains
switch to disconnect the mains supply of the device in the building installation and secure it against being
switched on again, e.g. by locking the switch.
Before any other cleaning or decontamination process other than that recommended by the manufacturer is
applied, the user has to check with the manufacturer that the planned process does not damage the device.

Centrifuges, rotors and accessories must not be cleaned in rinsing machines.

They may only be cleaned by hand and disinfected with liquids.

The water temperature must be between 20 – 25°C.

Only detergents/disinfectants may be used which:
 have a pH between 5 - 8
 do not contain caustic alkalis, peroxides, chlorine compounds, acids and alkaline solutions

In order to prevent appearances of corrosion through cleaning agents or disinfectants, the application guide from
the manufacturer of the cleaning agent or disinfectant are absolutely to be heeded.
35.1

Centrifuge (housing, lid and centrifuging chamber)

35.1.1

Surface cleaning and care


Clean the centrifuge housing and the centrifuging chamber regularly, using soap or a mild detergent and a damp
cloth if required. For one thing, this services purposes of hygiene, and it also prevents corrosion through
adhering impurities.

Ingredients of suitable detergents:
soap, anionic tensides, non-ionic tensides.

After using detergents, remove the detergent residue by wiping with a damp cloth.

The surfaces must be dried immediately after cleaning.

In the event of condensation water formation, dry the centrifugal chamber by wiping out with an absorbent cloth.

Lightly rub the rubber seal of the centrifuge chamber with talcum powder or a rubber care product after each
cleaning.

The centrifuging chamber is to be checked for damage once a year.
If damage is found which is relevant to safety, the centrifuge may no longer be put into operation. In this
case, notify Customer Service.
35.1.2

Surface disinfection


If infectious materials penetrates into the centrifugal chamber this is to be disinfected immediately.

Ingredients of suitable disinfectants:
ethanol, n-propanol, ethyl hexanol, anionic tensides, corrosion inhibitors.

After using disinfectants, remove the disinfectant residue by wiping with a damp cloth.

The surfaces must be dried immediately after disinfecting.
52/132
so that the motor-driven lid locking once again assumes