Herunterladen Inhalt Inhalt Diese Seite drucken

Centrifugation Of Materials With Higher Density; Rotor Identification; Emergency Release; Maintenance And Servicing - Hettich ROTOFIX 32 A Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für ROTOFIX 32 A:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

EN
17

Centrifugation of materials with higher density

The rotors are designed to centrifuge substances up to a maximum mean homogenous density of 1.2 kg/dm
rotating at the stated speed.
Denser substances must be centrifuged at lower speed.
The permissible speed can be calculated using the following formula:
Reduced
speed
(n
)
red
e.g.: RPM 4000, density 1.6 kg/dm
1.2
n
=
x
4000
=
3464
red
1.6
If in doubt you should obtain clarification from the manufacturer.
18

Rotor Identification

After every start of a centrifugation run the rotor utilised is identified.
After a rotor change, the drive switches off and the rotor code (rot xx) is displayed.
Press the key
. The last used centrifuge data will be displayed..
START
A further operation of the centrifuge is only possible after a single opening of the lid.
If, following a rotor change, the maximum speed of the rotor is less than the set speed, the speed is limited to
the maximum speed of the rotor.
19

Emergency release

The lid cannot be opened during power failure. An emergency release has to be executed by hand.
For emergency release disconnect the centrifuge from the mains.
Open the lid only during rotor standstill.
Only the plastic release pin provided may be used for emergency release.
See figure on page 2.
Switch off the mains switch (switch position "0").
Look through the window in the lid to be sure that the rotor has come to a standstill.
Insert the release pin (see scope of supply) horizontally into the hole (Fig. 1, A). Push the unlocking pin in until
the handle can be lifted when the pin is pressed down.
Open the lid.
20

Maintenance and servicing

Pull the mains plug before cleaning.
Before any other cleaning or decontamination process other than that recommended by the manufacturer is
applied, the user has to check with the manufacturer that the planned process does not damage the device.
Cleaning agents and disinfectants which lie in the pH range 5 – 8 are to be utilised. Alkaline cleaning agents with
a pH value > 8 are to be avoided.
In order to prevent appearances of corrosion through cleaning agents or disinfectants, the application guide from
the manufacturer of the cleaning agent or disinfectant are absolutely to be heeded.
20.1

Centrifuge

Clean the centrifuge housing and the centrifuging chamber regularly, using soap or a mild detergent and a damp
cloth if required. For one thing, this services purposes of hygiene, and it also prevents corrosion through
adhering impurities.
In the event of condensation water formation, dry the centrifugal chamber by wiping out with an absorbent cloth.
If infectious materials penetrates into the centrifugal chamber this is to be disinfected immediately. For surface
disinfection we recommend Bacillol® manufactured by Bode Chemie in Hamburg or Biocidal ZF
company WAK-Chemie Medical GmbH in Steinbach.
Lightly grease the rubber seal of the centrifugal chamber after every cleaning.
26/72
1.2
=
x
Rated
speed
Greater
density
3
RPM
3
when
TM
from the

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis