14.4

Run-up and run-down parameters

The run-up and run-down parameters that have been set are displayed.
x: 1-9 = run-up step, t = run-up time
y
x
y: R1-R9, B1-B9 = braking step, R0 = unbraked run-down, t = run-down time, n
energisation speed
PROFIL
14.4.1

Run-up step


Press the
key repeatedly until the parameter
inversely.

Set the desired step by means of the control knob
14.4.2

Run-up time


Press the
key repeatedly until the parameter

Set the desired value by means of the control knob
If a run-up time is set that is longer than the running time, the centrifugation run will end before the set speed is
reached.
14.4.3

Braking step


Press the
key repeatedly until the parameter
inversely.

Set the desired step by means of the control knob
B-braking steps can only be set for special rotors.
14.4.4

Run-down time

If a brake de-energisation speed has been set, it is not possible to set a run-down time.

Press the
key repeatedly until the parameter

Set the desired value by means of the control knob
14.4.5

Brake de-energisation speed


Press the
key repeatedly until the parameter n

Set the desired value by means of the control knob
14.5

Radius/Temperature

14.5.1

Radius


Press the
key repeatedly until the parameter r/mm is displayed and the input field is shown inversely.

Set the desired value by means of the control knob
The value of the RCF/RZB adapts automatically when the radius is changed; this is shown by the indication flashing.
14.5.2
Temperature

Press the
key repeatedly until the parameter T/°C is displayed and the input field is shown inversely.

Set the desired value by means of the control knob
14.6

Automatic buffer

The buffer comprises the program locations "----" and 90 to 99. Every time a centrifugation run is started, modified
centrifugation data are automatically stored in program location "----". The modified centrifugation data for the last 11
centrifugation runs are stored in the buffer and can be called up (see Chapter "Calling up programs").
46/132
or
is displayed. The input field will be displayed
t
1-9
.
is displayed. The input field will be displayed inversely.
min:sec
t
.
or
is displayed. The input field will be displayed
t
0-9
.
is displayed. The input field will be displayed inversely.
t
min:sec
.
(•)
/RPM is displayed. The input field will be displayed inversely.
.
.
.
(•)
= brake de-