31

Relative centrifugal force (RCF)

The relative centrifugal force (RCF) is given as a multiple of the acceleration of gravity (g). It is a unit-free value and
serves to compare the separation and sedimentation performance.
These values are calculated using the formula below:
RCF
RCF = relative centrifugal force
RPM = rotational speed (revolutions per minute)
r = centrifugal radius in mm = distance from the centre of the turning axis to the bottom of the centrifuge.
The relative centrifugal force (RCF) stands in relation to the revolutions per minute and the centrifugal
radius.
32

Centrifugation of materials or mixtures of materials with a density higher than 1.2 kg/dm

When centrifuging with maxim revolutions per minute the density of the materials or the material mixtures may not
3
exceed 1.2 kg/dm
.
The speed must be reduced for materials or mixtures of materials with a higher density.
The permissible speed can be calculated using the following formula:
Reduced
speed
(n
)
red
e.g.: maximum speed RPM 4000, density 1.6 kg/dm
1.2
kg/dm³
n
=
x
4000
red
1.6
kg/dm³
In the exceptional case that the maximum loading indicated on the hanger is exceeded, the speed must also be
reduced.
The permissible speed can be calculated using the following formula:
Reduced
speed
(n
)
red
e.g.: maximum speed RPM 4000, maximum load 300 g, actual load 350 g
300
g
n
=
x
4000
RPM
red
350
g
If in doubt you should obtain clarification from the manufacturer.
33

Rotor recognition

Rotor recognition is carried out after each start of the centrifugation run.
If the rotor has been changed, the centrifugation run is canceled after rotor recognition. The rotor code (R xx) and the
maximum speed (n-max=xxxxx) of the rotor are displayed.
A further operation of the centrifuge is only possible after a single opening of the lid.
If the maximum speed of the rotor being used is less than the set speed, the speed is limited to the rotor's
maximum speed.
2
RPM
r
1,118
1000
For more on the centrifugal radius see the chapter " Anhang/Appendix,
Rotoren und Zubehör/Rotors and accessories".
1.2
=
Greater
density
[kg/dm³]
3
RPM
=
3464
RPM
maximum
load
[g]
=
x
maximum
actual
load
[g]
=
3703
RPM
RCF
RPM
r
1,118
x
maximum
speed
[RPM]
speed
[RPM]