Set the sound level calibrator power switch to the 94dB or 114dB position,
depending on the desired level of calibration. Select the level closest to the
upper limit of the level range intended for measurement.
3.
Place the microphone of the sound level meter into the output of the sound
level calibrator.
4.
Switch on the sound level meter and check the reading on the display. The
sound level meter must me readjusted if there is a difference between the
display reading and the output of the sound level calibrator ( 94dB or 114dB)
5.
When the calibrator is not in use, please power off to save the battery
CAUTION!
Ambient sources of noise or vibration can cause a wrong calibration. Do the
calibration in quiet surrounding.
The low air pressure in high places has an effect on sound level measurements.
This have to take into consideration during the calibration.
Heigth
Pressure in
in m
mbar
amsl
0
1013
100
1000
200
990
300
978
400
966
500
955
600
944
700
932
800
921
900
910
1000
899
1100
889
1200
878
1300
868
1400
858
1500
847
set
set value at
value at
114dB
94dB
94,0
114,0
94,0
114,0
94,0
114,0
93,9
113,9
93,9
113,9
93,9
113,9
93,9
113,9
93,9
113,9
93,9
113,9
93,8
113,8
93,8
113,8
93,8
113,8
93,8
113,8
93,8
113,8
93,8
113,8
93,8
113,8
Heigth
Pressure in
in m
mbar
amsl
1600
837
1700
827
1800
816
1900
806
2000
796
2100
786
2200
777
2300
768
2400
758
2500
749
2600
739
2700
730
2800
721
2900
711
3000
702
9
set
set
value at
value at
94dB
114dB
93,7
103,7
93,7
103,7
93,7
103,7
93,7
103,7
93,7
103,7
93,7
103,7
93,6
103,6
93,6
103,6
93,6
103,6
93,6
103,6
93,6
103,6
93,6
103,6
93,6
103,6
93,5
103,5
93,5
103,5

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

12 05 99

Inhaltsverzeichnis