Herunterladen Inhalt Inhalt Diese Seite drucken

COSINUS ESCORT 3145A Bedienungsanleitung Seite 97

Inhaltsverzeichnis

Werbung

ESCORT 3145A / 3146A
CONFigure]
[:SCALar]
:RANGe? [
De
scription : Return the measurement range. Use @1 to select the primary display, and
@2 to select the secondary display.
Return the range value as Table 7-4.
Response :
CONF:RANG?
Example :
CONF:RANG? ,@2
Slow Rate
Function
DC Voltage
AC / AC+DC
Voltage
DC / AC /
AC+DC
Current
Resistance
Cosinus Computermesstechnik GmbH • Tel. 089-665594-0 • www.cosinus.de
[ ,@1] | ,@2 ]
; Query the primary display measurement range.
; Query the secondary display measurement range.
Return
Value
120mV
0.12
1.2V
1.2
12V
12
120V
120
1000V
1000
120mV
0.12
1.2V
1.2
12V
12
120V
120
750V
750
12mA
0.012
120mA
0.12
1200mA
1.2
12A
12
120Ω
120
1.2KΩ
1200
12KΩ
1.2E+4
120KΩ
1.2E+5
1.2MΩ
1.2E+6
12MΩ
1.2E+7
120MΩ
1.2E+8
Medium / Fast
Rate
400mV
4V
40V
400V
1000V
400mV
4V
40V
400V
750V
40mA
120mA
1200mA
12A
400Ω
4KΩ
40KΩ
400KΩ
4MΩ
40MΩ
300MΩ
Seite 7-95
Return
Value
0.4
4
40
400
1000
0.4
4
40
400
750
0.04
0.12
1.2
12
400
4000
4E+4
4E+5
4E+6
4E+7
3E+8

Quicklinks ausblenden: