7.5

Instructions of Command Sets

7.5.1 IEEE 488.2 Common Commands

*CLS
Description : Clear the Standard Event Status Register.
*ESE <NRf>
Description : Set the Standard Event Status Enable Register.
Parameter : <NRf> (0 to 255).
Example : *ESE 140; Enable bits 2 (QYE), 3 (DDE), and 7 (PON), and disable all
the other bits.
*ESE?
Description : Query the Standard Event Status Enable Register.
Response : (Integer) Decimal equivalent of the register byte. Range is 0 to 255.
*ESR?
Description : Query the Standard Event Status Enable Register and clears the
register.
Response : (Integer) Decimal equivalent of the register byte.
Example : *ESR? ; Return: '32'if bit 5 (CME) is set (1) and the rest of the bits are
reset (0).
*IDN?
Description : Query the Meter identification.
Response : Return the following information : Firmware version and Model of the
meter ; Firmware version and Model of the GPIB card.
Example : *IDN? ; Return: 'v1.31,6 ; v1.1,08'represents the Model-A meter installed
with firmware version "v1.31" and the Model 08 GPIB card installed with
firmware "1.1".
*OPC
Set the Operation Complete bit in the Standard Event Status
Description :
Register when all pending device operations are complete.
*OPC?
Description : This command pauses program execution until all operations are
completed.
Response : Return a '1' after all pending operations are complete.
*RST
Description : Place the meter in its power-on-reset state. This command does not
affect:
Cosinus Computermesstechnik GmbH • Tel. 089-665594-0 • www.cosinus.de