Probe rod / rope without process connection
Tool list
If necessary use a multigrip piers for the disassembly of probe.
Do not use any tools for re-assembly, the coating could be damaged.
2 Exchange of probe rod / rope
a
Sonde FMP52_1
3 Re-assembly
Re-assembly is carried out in reverse order, unless otherwise instructed.
4 Exchange of coax probe - FMP55 only
a
Sonde_FMP55_1
5 Re-assembly
Re-assembly is carried out in reverse
order.
Please take care to the position of the
spacers, like shown beside.
6 Re-calibration of the instrument
Install the instrument into the vessel.
If the probe length is identical with the original length of the probe, a probe calibration is not required.
If the probe length or probe type is different, perform a new level calibration according to the description in the Operating Manual.
Additionally it is necessary to order a new typeplate and fix it on the instrument!
Endress+Hauser
b
Sonde_FMP52_2
b
Sonde_FMP55_2
Sonde_FMP55_4
c
Sonde_FMP55_3
Levelflex FMP52, FMP55
2