to Type Examination Certificate No. FM15ATEX0008X
13
Description of Equipment or Protective System:
The PMA.Evolution EVO1Yab-c and LAB1Yab-c are designed for the measurement of weight in the
hazardous location. The PMA.Evolution contains the processing circuitry and has a touch enable LCD
display with backlight facility. Power and data are provided via a USB connection.
Uin = +15Vdc / +5Vdc
Um = 20 Vdc.
An optional power supply YPS04-Y.. (KEMA 09ATEX0121X) can be used to provide power if
necessary.
The PMA.Evolution is available in two versions EVO1Yab-c and LAB1Yab-c.
PMA.Evolution EVO1Yab-c
a = Up to three letters and/or numbers or blank (not critical to safety)
b = Up to two letters and/or numbers or blank (not critical to safety)
c = Up to five letters and/or numbers or blank (not critical to safety)
PMA.Evolution LAB1Yab-c
a = Up to three letters and/or numbers or blank (not critical to safety)
b = Up to two letters and/or numbers or blank (not critical to safety)
c = Up to five letters and/or numbers or blank (not critical to safety)
14
Specific Conditions of Use:
1. The enclosure of the PMA.Evolution EVO1Yab-c and LAB1Yab-c are non-metallic and shall not be
used where UV light or radiation may impinge on the enclosure.
2. The PMA.Evolution EVO1Yab-c and LAB1Yab-c shall be installed in such a way that it is protected
against the entry of solid foreign objects or water capable of impairing the safety of the apparatus.
Reduce the risk of mechanical damage to a minimum.
3. The enclosure of the PMA.Evolution EVO1Yab-c and LAB1Yab-c is non-conducting and may
generate an ignition-capable level of electrostatic charges under certain extreme conditions. The user
shall ensure that the equipment is not installed in a location where it may be subjected to external
conditions that might cause a build-up of electrostatic charges on non-conducting surfaces, additionally,
cleaning of the equipment should be done only with a damp cloth.
15
Essential Health and Safety Requirements:
The relevant EHSRs that have not been addressed by the standards listed in this certificate have been identified
and assessed in the confidential report identified in item 8.
16
Test and Assessment Procedure and Conditions:
This Type Examination Certificate is the result of testing of a sample of the product submitted, in
accordance with the provisions of the relevant specific standard(s), and assessment of supporting
documentation. It does not imply an assessment of the whole production.
Whilst this certificate may be used in support of a manufacturer's claim for CE Marking, FM Approvals Ltd
accepts no responsibility for the compliance of the equipment against all applicable Directives in all
applications.
This Certificate has been issued in accordance with FM Approvals Ltd's ATEX Certification Scheme.
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals Ltd. 1 Windsor Dials, Windsor, Berkshire, UK. SL4 1RS
T: +44 (0) 1753 750 000 F: +44 (0) 1753 868 700 E-mail:
F ATEX 029 (Apr/16)
SCHEDULE
atex@fmapprovals.com www.fmapprovals.com