General specifications apply for the temperature range 5°C to 40°C. Accuracy specifications apply
for the temperature range 18°C to 28°C after 1 hour warm-up with no load and calibration at 23°C.
Typical specifications are determined by design and are not guaranteed.
OUTPUTS
Voltage/Current Ranges:
PL 068
PL 155
PL 303
PL 601
PL303QMD &
PL303QMT
Note: In manual operation, actual maxima for voltage and current are typically 1% greater than the
figures given above.
Voltage Setting:
Current Setting:
Configuration Selection:
(PL303QMD & QMT only)
Voltage Span Control
(V-Span)
Settings Lock
(S-Lock)
Output Mode:
Output Switch:
Output Terminals:
Transient Response:
Voltage Programming
Speed (PL-P models):
6V 8A
6V 800mA
15V 5A
15V 500mA
30V 3A
30V 500mA
60V 1·5A
60V 500mA
2
0V to 6V/1mA to 8000mA; 0V to 6V/0.1mA to 800mA
0V to 15V/1mA to 5000mA; 0V to 15V/0.1mA to 500mA
0V to 30V/1mA to 3000mA; 0V to 30V/0.1mA to 500mA
0V to 60V/1mA to 1500mA; 0V to 60V/0.1mA to 500mA
0V to 30V/1mA to 3000mA; 0V to 30V/0.1mA to 500mA (each 30V output);
0V to 30V/2mA to 6000mA; 0V to 30V/0.2mA to 1000mA (parallel mode).
& 0V to 6V/1mA to 8000mA; 0V to 6V/0.1mA to 800mA (PL303QMT O/P3)
By coarse and fine controls.
By single logarithmic control.
Independent, True parallel, Isolated Tracking & Isolated Ratio Tracking
modes via front panel rotary switch.
The voltage adjustment range can be controlled by digital setting of the end-
stop values of the coarse voltage control to any desired values.
The range for Vmax is 0.1V to 6V/15V/30V/60V depending on model.
The range for Vmin is 0 to (Vmax – 0.1V).
Voltage and current settings can be locked by a single button press.
Lock accuracy is equal to meter accuracy (see Meter Specification)
Constant voltage or constant current with automatic cross-over.
CC indicator lit in constant current mode.
Electronic, non-isolating.
Preset voltage and current limit displayed when Output is off.
Output rise time with no load <15ms.
Universal 4mm safety binding posts on 19mm (0·75") spacing for Output;
screwless terminals for Sense.
<50µs to within 50mV of set level for a 5% to 95% load change.
Maximum time required for output to settle within 1% of its total excursion (for
resistive load). Excludes command processing time.
Full Load
Up
20ms
Up
20ms
Up
45ms
Up
45ms
Up
45ms
Up
45ms
Up
45ms
Up
70ms
No Load
5ms
Down
5ms
Down
40ms
Down
40ms
Down
40ms
Down
40ms
Down
40ms
Down
40ms
Down

Specification

Full Load
No Load
5ms
20ms
6ms
60ms
20ms
50ms
50ms
110ms
80ms
80ms
100ms
100ms
150ms
150ms
300ms
300ms