Technical data

11
Technical data
Measuring principle
Measuring range
Breath
alcohol
concentration
Blood alcohol
concentration
Display range
The range of indication and the measuring range are identical.
If the result is exceeding the measuring range the display will show
three arrows "↑ ↑" as overflow indication.
34
Electrochemical Sensor
0.00 to 2.5 mg/L
0 to 2500 µg/L
0.000 to 0.500 g/210L
0 to 250 µg/100mL
0 to 250 µg%
(Ethanol mass per breath volume at
34 °C and 1013 hPa.)
0.00 to 5.00 ‰
0.000 to 0.500 %
0.00 to 5.00 g/L
0 to 500 mg/100 mL
0 to 500 mg%
0.00 to 5.00 mg/mL
(Ethanol mass per blood volume resp.
o
mass at 20
C and 1013 hPa.
By giving this data a conversion factor
of 2100 between breath and blood
alcohol concentrations is used.)
Ambient conditions
For operation
For storage
Sample conditions
minimal blowing time
minimal volume
Measurement accuracy
Reproducibility with an ethanol standard
up to 0.50 mg/L
≥ 0.50 mg/L
up to 1.00 ‰
≥ 1.00 ‰
Values for other measuring units can be calculated according to
table on page 34.
Sensitivity drift
o
–5 to +50 °C (23 to 122
F)
10 to 100 % r.h.
(non condensing)
600 to 1400 hPa
o
–40 to +65 °C (–40 to 149
F)
selectable
2 s default
selectable
1.2 L default
0.008 mg/L
1.7 % of measurement volume
the higher volume is valid
0.017 ‰
1.7 % of measurement volume
the higher volume is valid
typically 0.6 % of meas. value/month