A Information about the modified device
Relevant documentation
Not only the already existing documentation, but also the documentation delivered with the transmitter will be relevant additionally,
because of the conversion.
For the sensor the documentation for Promag 50, 53 is relevant.
For the transmitter the documentation for Promag 55 is relevant.
Changes and restrictions conditioned by conversion
• Accuracy
After conversion of the device an additional measured error of ± 0,2% to the measured value must be calculated.
B Preparatory operations: Save device data
• Conversion Promag 50
Select important parameter out of the application (e.g. with the aid of the Quick Setup "Commisioning" see Operating Instructions) and
fill them into the "list of parameter", chapter H. After Conversion enter the noted parameter into the new transmitter.
• Conversion Promag 53
Save the parameters of the transmitter to the T-DAT using the "T-DAT SAVE/ LOAD" function".
Replace the T-DAT into the transmitter after conversion.
C Overview data storage (Histo-ROM) and F-CHIP
Data storage
S-DAT
T-DAT
F-CHIP (Optional)
Endress + Hauser
Conversion transmitter Promag 50/53 W, P, H to Promag 55 remote version
Colour
blue
black
red
Promag 50 to Promag 55
Re-assembly
New
New
Promag 53 to Promag 55
Re-assembly
Re-assembly
New