Technical data

11
Technical data
Measurement principle
Measuring range
Breath alcohol concentration 0.00 to 2.50 mg/L
Blood alcohol concentration 0.00 to 5.00‰
Indication range
The display is identical to the measuring range.
If the measurement result is out of range, the display will show
three arrows "↑ ↑ ↑" indicating that the measuring range has been
exceeded.
34
Electrochemical sensor
(Ethanol mass per breath volume at
34 °C and 1013 hPa.)
(Ethanol mass per blood volume
and/or mass at 20 °C and
1013 hPa.
For the above data, a conversion
factor of 2100 is used for breath and
blood alcohol concentration data
(works setting, country-specific
conversion factors can be
configured).
Ambient conditions
During operation
During storage
Taking samples
Breath volume
Breath flow
Duration of breath
Measurement accuracy
Standard deviation of the repeatability with ethyl alcohol standard
up to 0.50 mg/L
≥ 0.50 mg/L
up to 1.00 ‰
≥ 1,00 ‰
Sensitivity drift
–5 to +50 °C
10 to 100 % r.h.
(non-condensing)
600 to 1400 hPa
–40 to +70 °C
1,2 l (pre-set)
>10 l/min (pre-set)
2 s (pre-set)
0,008 mg/l
1.7 % of the measured value
0,017 ‰ or
1.7% of the measured value
whichever value is higher
typically 0.6% of the measured
value/month
Dräger Alcotest 6820