TROUBLESHOOTING GUIDE

Problems
A beep is heard and a
message appears on the
display to check the
belt's adhesion to the skin.
The display works, but
the belt does not
generate stimulation.
Contractions are very
weak even when the
intensity is high, and
discomfort is felt.
The display does not
turn on.
Issues different from
those listed above.
Possible Causes
The belt is not properly
adhered to the skin.
The control device is
not correctly positioned
on the belt.
The conductive pads and
skin are not sufficiently
moistened.
The presence of dust
deposits or oxidation on
the metal contact pins
between the belt and
the device may prevent
electrical contact.
The intensity may
be set too low.
The conductive pads
and skin may not be
sufficiently hydrated.
The battery is discharged. Recharge the battery.
The belt or one of its
components may be
defective.
47
Solutions
Tighten the belt to ensure
optimal contact with
the skin.
Ensure that the control
device is properly
attached to the belt in
the correct direction
using the magnets.
Moisten both the
conductive pads and the
skin to improve conduction.
Use a stiff brush,
preferably metallic, to
carefully clean the metal
buttons, especially in the
recessed area where
contact occurs. Then,
blow to remove any
remaining debris.
Back
Front of the
of the device
belt
Press the "+" button to
increase the intensity and
check that you are
adjusting the level of the
currently used channel.
Moisten both the
conductive pads and
the skin to improve
conduction.
Contact TESMED
customer support at
info@tesmed.com.