Non-verified models:
Switching the weight unit
Restoring factory settings
(rESEt)
54 • seca 675/seca 674
3. Select a filtering level.
4. Confirm your selection.
ð The setting is adopted.
ð The menu closes.
You can select the unit in which weight is to be displayed.
CAUTION!
(UnIt)
Patient hazard
To prevent misinterpretations, measuring results for medical purposes must only
be displayed and used in SI units (kilograms/grams, meters/centimeters). Some
devices have the option of displaying measuring results in different units. This is
purely an additional function.
► Only use the measuring results in SI units.
► You as the user take sole responsibility for the use of measuring results in
non-SI units.
1. From the menu, select the
2. Confirm your selection.
ð The current setting is displayed.
3. Select the unit in which weight is to be displayed:
– Kilograms/grams (
– Pounds (
– Stones (
4. Confirm your selection.
ð The setting is adopted.
ð The menu closes.
You can restore the factory settings for the functions below:
Autohold (
AHOLd
Acoustic signal (
Acoustic signal (
Filtering (
)
FIL
Autoclear (
ACLr
Pre-tare (
)
Pt
Height for Body Mass Index (
Unit, weight (
UnIt
1. From the menu, select the
2. Confirm your selection.
ð The menu closes.
3. Switch off the scale.
ð Factory settings are restored and are available when the scale is switched on.
menu item.
UnIt
gr
)
)
lbS
)
sts
Function
)
PrESS
)
)
HOLd
)
)
bMi
)
menu item.
rESEt
Factory setting
model-depen-
dent
OFF
On
0
model-depen-
dent
0 kg
170 cm
(
)
kg
gr