Explanation Of Symbols; Consumables /Accessories; Spare Parts - Vitalograph 2120 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für 2120:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 12
C
D
LEANING AND
ISINFECTING THE
V
2120
ITALOGRAPH
Definitions of cleaning and disinfection are as defined in
"Sterilization, Disinfection and Cleaning of Medical
Equipment: Guidance on Decontamination from the
Microbiology Committee to Department of Health Medical
Devices Directorate, 1996"
Recommendations for chemical disinfectants are derived
from the PHLS publication "Chemical Disinfection In
Hospitals" 1993.
* May be autoclaved dry or wet; up to 134ºC; up to 20 mins;
up to 50 cycles.
Part
Material
Case
A.B.S.
exterior
Contact
Tin plated
pins &
brass
sockets
Flowhead
T.P.X.
cone*
Flow
Acetal
conditioning
and
mesh
Polyester
End cap*
T.P.X.
Fleisch
Anodized
element*
Aluminium
&
stainless
steel
O-Rings*
Viton
* Will also withstand autoclaving at 134°C for 3 minutes.
Clean each separate part of the flowhead by washing in a
mild detergent to remove particulate contamination. To
clean the Fleisch element, swill vigorously in water with
mild detergent. Do not attempt to "rub" or "scrub" at
capillaries. If the flow conditioning meshes appear dirty or
blocked they should be thrown away and replaced.
Rinse with clean water.
Disinfect by immersion in sodium dichloroisocyanurate
solution at 1000 ppm concentration of free chlorine for 15
minutes. Prepare disinfectant solution as directed in the
manufacturer's guidelines.
Rinse with hot water to aid drying.
End Cap
Leave to dry completely before reassembling. Drying the
Fleisch element assembly may require placing it in a warm
place overnight. A drying cabinet is ideal, alternatively
another heat source could be used.
Wiping with a 70% Isopropyl Alcohol impregnated cloth
provides a suitable form of cleaning and low-level
disinfection for the case exterior, display, and keypads.
Repeat at least weekly to prevent build-up of grime from
normal handling and use.
Always follow the safety guidelines given by the manufacturer
of cleaning and disinfectant chemicals or equipment.
10
Recommended
Disinfectants
Wiping with a 70%
Isopropyl Alcohol
impregnated cloth
provides a suitable form of
cleaning and low-level
disinfection.
Disinfect by immersion in
sodium
dichloroisocyanurate
solution at 1000 ppm
concentration of free
chlorine for 15 minutes.
(see recommended
cleaning/disinfection
method for the
Vitalograph 2120)
Flow Conditioning Mesh
Fleisch Element
2120 Unit
Flow
Conditioning
Mesh
E
XPLANATION OF SYMBOLS
k
Type BF equipment
Class II
VA
Power rating
V
Voltage DC
h
Attention (reference relevant section in manual)
Other Labels
C
ONSUMABLES
Cat. No.
Description
20242
SafeTway Mouthpieces (200)
20303
Noseclips (10)
20408
1-L Precision Syringe
36020
3-L Precision Syringe
28350
BVF (50)
42084
Flow Conditioning Meshes (10)
74075
Spirotrac to Cradle Connectivity Kit
2120505
2120 Intelligent Printer Cable
S
P
PARE
ARTS
Cat. No.
Description
31297SPR
Power Cord - UK
31375SPR
Power Cord - USA
31376SPR
–Power Cord - European
62019SPR
Flowhead Cone
62006SPR
Flowhead End Cap
66408SPR
Large Carrying Pouch
20979
Large Instrument Briefcase with Foams
74076
2120 to Cradle Coily Cable
76062
9 Way to 9 Way Serial Cable
21001204
PowerSAFE
2120060
2120 Serial and power lead cable
Cone
Serial connector
Power input connector
On (I) / Off (O) switch
Power indicator
Docking indicator
/A
CCESSORIES