Illumination tester
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Characteristics SECULIFE
Measuring Quantity
Illumination
Luminance
with luminance
attachment
for cd/m² or fL
Most Important Error Limits SECULIFE
Characteristics
V()-Adaptation (f
)
1
True Cosine Evaluation (f
Linearity (f
)
3
Adjustment Error (f
)
11
Overall Error (f
)
ges
GMC-I Messtechnik GmbH
IT
Measuring Range
in Lux (lx)
I
0.01...
II
0.1...
III
1...
IV
10...
V
100... 199 000
in Candela/m² (cd/m²)
I
0.1...
II
1...
III
10...
IV
100...
V
1000... 1 999 000
IT
Admissible Error according to
DIN 5032 Class B
)
2
Measuring Range
in footcandle (fc)
19.99
0.001...
199.9
0.01...
1 999
0.1...
19 900
1...
10...
in footlambert (fL)
199.9
0.01...
1 999
0.1...
19 990
1...
199 900
10...
100...
6%
3%
2%
1%
10%
Resolution
in lx
1 999
0.01
19.99
0.1
199.9
1
1 999
10
19 990
100
in cd/m²
19.99
0.1
199.9
1
1 999
10
19 990
100
199 900
1000
Typical Error
3.0%
2.0%
1.0%
0.8%
~ 8.0%
Resolution
in fc
0.001
0.01
0.1
1
10
in fL
0.01
0.1
1
10
100
17