Cleaning Recommendations

When cleaning the probe, take care not to scratch or damage the sensing surface or the
platinized electrode surfaces.
Contaminant
Cleanin
Solution
Water soluble substances
Deionized
water
Grease
& Oil
Warm water and household
detergent
Heavy grease & Oil
Alcohol
Lime and hydroxide
10%
acetic
acid
coatings
Please Note: Since the EC500 does not have a refillable reference electrolyte chamber, it is
important not to soak the electrode in the above solutions for more than the recommended
times. To do so may cause a reference potential shift, which will cause degradation in
pefformance
or failure.
12
Instructions
Soak or scrub with a soft
brush. Recondition
in 4 or 7
buffer for I hour.
Soak
or scrub
with
a soft
brush, maximum of 10
minutes. Rinse thoroughly
with Dl water, recondition
in
4 or 7 buffer
for 1 hour.
Maximum of 5 minute soak,
scrub with a soft brush.
Rinse thoroughly with Dl
water, recondition in 4 or 7
buffer for 1 hour.
Soak until coating dissolved,
maximum
of 5 minutes.
Rinse thoroughly
with DI
water. recondition
in 4 or 7
buffer
for I hour.
EC500-EU
v2.o
5/05
Troubleshooting
Possible
Cause
Problem
Readingis frozen
unit is in "HOLD" mode
Batteries
are low
"BAT"
messa
e
Low pH stope
Unit will not calibrate in pH
Unit will not calibrate in pH
Clogged or contaminated
reference
•unction
Unit will not calibrate in pH
Damaged or worn out
sensin
membrane
Contaminated
Unit will not calibrate
in
H
Unit will not calibrate
in
Contaminated conductivity
conductivity
mode
standards
Unit will not calibrate
in
Dirty probe
conductivi
mode
IJnit will not calibrate
in
Damaged conductivity probe
conductivi
mode
Trapped air bubbles
Unit will not
calibrate
in
conductivi
mode
Unit wit( not turn on
Battéries
are tow or dead
Batteries
installed
Unit
will not turn
on
incorrect
Olari
"RENEW" message
pH sensor needs
recalibration
"RENEW" message
pH sensor slope has fallen
below
acce
table limits
Unit will not respond to any
Internal
fault
key presses
Copyright
0 2005 Extoch Instruments
All rights reserved
including
the right of reproduction
wvm.extech.com
Action
Press MODE/HOLD key to
exit "HOLD"
mode
Re lace batteries
Replace electrode, see
reorder
information
Clean junction (see
cleanin
instructions
Replace electrode. see
reorder
information
Use fresh
buffers
H buffers
use fresh
standards
Clean conductivity
probe
See
cleanin
instructions
Replace electrode, see
reorder
information
Tap or stir to release air
bubbles
Re
lace batteries
with
Replace batteries,
observe
Olari
Recalibrate
unit
use
fresh
buffers
Replace electrode, see
reorder
information
Perform
hard reboot:
Remove
batteries,
hold
ONIOFF
switch
down for 5
seconds,
re lace batteries