4
CLEANING AND DISINFECTION
Regularly clean all product components and the accessories
used, if they are visibly dirty immediately after use.
Disinfect all individual parts at least once a week, in the event
of an acute infection, once a day.
4.1 Preparation
• Take the metered dose inhaler out of the connecting ring on
the inhalation chamber and close it with the protective cap
provided for this purpose.
• If applicable, detach the mask and the mask connector from
the mouthpiece.
• Disconnect the mouthpiece from the inhalation chamber.
• If applicable, remove the one-handed operation aid from the
inhalation chamber.
Info: The protective cap must be
open.
4.2 Cleaning
By hand
• Place all disassembled components in warm tap water with
a little dishwashing liquid for at least 5 min.
• Rinse all parts thoroughly in running water.
• You can remove excess water more quickly by shaking all
parts.