Herunterladen Inhalt Inhalt Diese Seite drucken

B2.4A; B2.4B; B2.4C - Flaem Rf9 Bedienungsanleitung

Inhaltsverzeichnis
2. Position the valve (B2.4.b) on the particle size regulator (B2.4.a) with the flat side down and align-
ing the hole with the position ribs, and then screw the selector (B2.4.c) onto the threaded pin of the
regulator as far as it will go. Turn the assembled regulator to ensure that it rotates freely and that the
selector switch rises and falls between positions.

B2.4b

B2.4a

3. Insert and press the previously assembled par-
ticle size regulator (B2.4) all the way in, keeping
it horizontal, into the upper part (B2.3), matching
the selector bulkheads (B2.4.c) with the recess in
the upper part (B2.3).
4.
Insert
the
assembled
(B2.4+B2.3+B2.2) into the lower part (B2.1) and
engage it by turning it clockwise.
HOW TO USE THE "RF9" AMPOULE WITH PARTICLE SIZE REGULATOR AND VALVE SYSTEM
It is professional, fast and suitable for the administration of all types of drugs, even in patients with
chronic diseases. The valve system optimises the delivery of the nebulised drug to the patient and
limits its dispersion in the environment. In addition, thanks to the particle size regulator, it is possible
to select the size of the nebulised particles according to the area of the respiratory system for which
the therapy is intended. With the 4 possible modes, the ampoule can deliver particles of different
sizes.
B2.3
a
upper
part
21

B2.4c

B2.4a
+
B2.4b
B2.4c
B2.4
b
B2.4+B2.3+B2.2
B2.1
Inhaltsverzeichnis