6.3 Calibration/Adjustment

The testo 265 should be regularly checked, and if neces-
sary adjusted, with the help of the Testo reference oil.
Preparing a sensor calibration / adjustment
The testo reference oil (0554 2650) is required to carry out
a sensor calibration / adjustment.
Warm the reference oil to room temperature (approx.
1
20-25°C) in a water bath prior to calibration / adjust-
ment.
Clean the sensor and the protective cage before
2
adjustment/calibration, in order to avoid transferring
contaminated oil into the reference oil, see
Cleaning the instrument
Carry out calibration / adjustment, see
3
instrument
Because of the high temperature fluctuations to which the
sensor is exposed in measurement operation, possible
contamination etc., we recommend an annual check by
Testo customer service.
We recommend the annual renewal of a calibration
certificate for measurements as part of a quality assurance
concept (e.g. ISO 9000).
Unlock protective cage using
1
bayonet nut
from probe pipe
Clean instrument parts, rinse off
2
with water and wipe dry.
Reattach protective cage to probe
3
pipe
.
.