Appendix 1 Guidance and Manufacturer Declaration Tables
Guidance and manufacturer´s declaration- electromagnetic emission
The Model PG-800B11 is intended for use in the electromagnetic environment specified below. The custo-
mer or the user of the Model PG-800B11 should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions CISPR 11 Group 1
RF emissions CISPR 11 Class B
Harmonic
emissions
A
IEC 61000-3-2
Voltage fluctuations /
Complied
flicker emissions IEC
61000-3-3
Guidance and manufacturer´s declaration – electromagnetic immunity
The Model PG-800B11 is intended for use in the electromagnetic environment specified below. The custo-
mer or the user of the Model PG-800B11 should assure that it is used in such an environment.
Immunity test
IEC 60601 Test level
Electrostatic discharge
± 8 kV contact
(ESD)
±2 kV, ±4 kV, ±8 kV, ±15
IEC 61000-4-2
kV air
Electrostatic transient/
± 2 kV for power supply
burs
lines 100 kHz repetition
frequency ±1 kV for in-
IEC 61000-4-2
put/output lines
Electromagnetic environment – guidance
The Model PG-800B11 uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The Model PG-800B11 is suitable for use in all establishments, in-
cluding domestic establishments and those directly connected to
the public low-voltage power supply network that supplies buil-
dings used for domestic purposes.
Compliance level
± 8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15
kV air
± 2 kV for power supply
lines 100 kHz repetition
frequency ±1 kV for in-
put/output lines
Electromagnetic
environ-
ment -guidance
Floors should be wood, con-
crete or ceramic tile. If floors are
covered with synthetic materi-
al, the relative humidity should
be at least 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
EN
41