Error codes and countermeasures
Error codes are indicated by a single letter in the lower right corner of the display. However, these only
occur in the menus M00, M01, M02, M03, M90 and M08. The following chart shows the error codes
and countermeasures.
Error
code
R
I
J
H
Q
F
1
2
3
K
Message in window
M08
System Normal
Detect No Signal
Hardware Error
PoorSig Detected
Frequ OutputOver
System RAM Error
Date Time Error
CPU or IRQ Error
ROM Parity Error
Adjusting Gain
Empty pipe
© PCE Instruments
34
Reason
No error
(1) No signal
(2) Sensors installed
improperly
(3) Too much fouling
(4) Liner too thick
(5) Sensor cable not
properly
connected
Hardware problem
(1) Poor signal
(2) Sensors installed
improperly
(3) Too much fouling
or contamination
(4) Liner too thick
(5) Sensor cable not
properly
connected
The output frequency
is outside the
permitted range
(1) Temporary
Problems with the
RAM or RTC
(2) Permanent
Problems with the
hardware
The device is
currently re-setting
the signal gain; the
number indicates the
current work
progress
(1) No liquid
in the pipeline
(2) Setting error
in menu M29
Countermeasure
- - -
(1) Change
measuring
location
(2) Clean measuring
location
(3) Check the cables
Contact PCE
Instruments
(1) Change
measuring
location
(2) Clean measuring
location
(3) Check the cables
(4) Check the
contact gel
Check the values in
the windows M67,
M68 and M69. Enter
higher values in
window M69.
(1) Restart the device
(2) Contact PCE
Instruments
- - -
(1) Choose a pipe
that contains
liquid
(2) Enter "0" in
window M29.