Herunterladen Diese Seite drucken

struers LaboPol-21 Gebrauchsanweisung Seite 29

Werbung

1
2
L1
A
MUST BE SHORT CIRCUIT
PROTECTED WITH EXTERNAL
FUSES ACCORDING TO LOCAL
L2
REGULATIONS
F = 16Amp MAXIMUM
PE
B
FROM
FT2
LaboForce-3
FT2
THERMAL PROTECTION M2
PE
Comment:
Use Dummy for LaboForce-1:
C
ALL WIRINGS 0,75mm², EXCEPT OTHERWISE MARKED.
D
COLOR CODES:
BK = BLACK
BN = BROWN
RD = RED
YE = YELLOW
GN = GREEN
BL = BLUE
GY = GREY
WH = WHITE
1
2
3
X1
QS1
2x5AMP
W1
F1
4
BK1
2
BK2
F2
PE
YE/GN
3x1mm²
POWER ON
REAR
PLATE
XPE1
X3
W3
1
3
2
3x0,75mm²
REAR
PLATE
XPE2
YE/GN
YE/GN
XPE5
XPE6
XPE4
BOTTOM
FRONT
BOTTOM
PLATE
PLATE
FRAME
3
4
5
5
3
3
S2
START
6
4
4
BK3
1
STOP
S1
2
BK4
2b
SELECT
S3
LaboForce-1/-3
W4
BK5
2
X5
BK5
BK2
W4
T1
U6
t>155C
2 ~
THERMAL
PROTECTION
M1
T2
W4
BK4
YE/GN
A2
K1
XPE3
A1
BK2
XPE3
REAR
PLATE
4
5
6
1
K1
2
SELECT
BK9
X2
LaboForce-1/-3
1
1a
BK10
L
S3
BK11
N
PE
YE/GN
3x1mm2
XPE2
REAR
PLATE
5x1mm²
BK1
U1
*
C1
M
U5
180W
M1
W4
MAINS SUPPLY:
*
C1
100V/50Hz.
50µF
100V/60Hz.
40µF
110-120V/60Hz.
30µF
220-240V/50Hz.
10µF
Thursday, September 13, 2001
220-240V/60Hz.
8µF
Monday, December 11, 2000
6
7
8
A
W2
POWER
L1
TO
L2
B
LaboForce-1/-3
PE
C
D
STRUERS A/S
VALHOEJSALLE 176
DK-2610 ROEDOVRE
DENMARK
PHONE:+45 3670 3500
CIRCUIT DIAGRAM
LaboPol-21
Size CAGE CODE
DWG NO
15293100
Rev
A2
SCALE
Sheet
of
1
1
DEM/DEM
7
8

Werbung

loading

Diese Anleitung auch für:

Labopol-25