Hettich HAEMATOKRIT 200 Gebrauchsanweisung Seite 79

Vorschau ausblenden Andere Handbücher für HAEMATOKRIT 200:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Fault description
no display
IMBALANCE
MAINS INTER 11, MAINS
INTERRUPT
TACHO - ERROR 1, 2
LID ERROR 4.1 - 4.127
OVER SPEED 5
VERSION-ERROR 12
UNDER SPEED 13
CTRL-ERROR 25.1-25.2
CRC ERROR 27.1
COM ERROR 31-36
FC ERROR 60, 61.1-61.21,
61.64-61.142
FC ERROR 61.23
TACHO ERR 61.22
FC ERROR 61.153
The left
half of the display lights up.
9.2
Performing a MAINS RESET
AB1801en
Cause
No power. Mains input fuses
defective.
The rotor is unevenly loaded.
Loss of mains power during
the centrifugation run. The
centrifugation run was not
completed.
Speed pulse failure.
Lid lock error.
Overspeed.
Wrong centrifuge model
detected. Error/defect in elec-
tronics.
Underspeed.
Error/defect in electronics.
Error/defect in electronics.
Error/defect in electronics.
Error/defect in electronics.
Speed measurement error.
Speed measurement error.
Error/defect in electronics.
-
1.
Set the mains switch to
Rev.: 10 / 11.2023
Remedy
Check the supply voltage.
Check mains input fuses.
The mains switch is in switch position
[I]
Open the lid.
Check the loading of the rotor.
Repeat the centrifugation run.
Open the lid.
[START/PULSE] button.
Press the
If required: Repeat the centrifugation
run.
Perform a MAINS RESET.
Perform a MAINS RESET.
Perform a MAINS RESET.
Perform a MAINS RESET.
Perform a MAINS RESET.
Perform a MAINS RESET.
Perform a MAINS RESET.
Perform a MAINS RESET.
Perform a MAINS RESET.
Do not switch off the device while
'Rotation' is displayed.
Perform a MAINS RESET if
locked' is displayed.
Do not switch off the device while
'Rotation' is displayed.
Perform a MAINS RESET if
locked' is displayed.
Perform a MAINS RESET.
Open the lid.
Check the loading of the rotor.
Repeat the centrifugation run.
Notify customer service.
[0] .
Troubleshooting
'Lid
'Lid
35 / 39

Werbung

loading

Diese Anleitung auch für:

18011801-01