Herunterladen Diese Seite drucken

Hettich HAEMATOKRIT 200 Gebrauchsanweisung Seite 73

Vorschau ausblenden Andere Handbücher für HAEMATOKRIT 200:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

7.2.3.2
Setting an audible signal
7.2.4
Visual signal
Switching on and off
AB1801en
after a problem occurs in the 2 s interval.
after completion of the centrifugation run and rotor standstill in the 30 s
interval.
Opening the lid or pressing any button stops the audible signal.
1.
Press and hold the
'*MACHINE MENU*' is displayed after 8 seconds.
��
[SELECT] button repeatedly until '-> Settings' is displayed.
2.
Press the
3.
Press the
[START/PULSE] button.
'End beep = on' or 'End beep = off' is displayed.
��
[t] buttons to set 'off' or 'on' .
4.
Use the
off: Audible signal after completion of the centrifugation run is disa-
bled.
on: Audible signal after completion of the centrifugation run is ena-
bled.
[SELECT] button.
5.
Press the
'Error beep = on' or 'Error beep = off' is displayed.
��
[t] buttons to set 'off' or 'on' .
6.
Use the
off: Audible signal after the occurrence of a malfunction is disabled.
on: Audible signal after the occurrence of a malfunction is enabled.
[SELECT] button.
7.
Press the
'Beep volume = min' , 'Beep volume = mid' or 'Beep volume =
��
max' is displayed.
[t] buttons to set 'min' , 'mid' or 'max' .
8.
Use the
min: The volume of the audible signal is set to low.
mid: The volume of the audible signal is set to medium.
Max: The volume of the audible signal is set to loud.
[START/PULSE] button.
9.
Press the
�� The setting is stored.
'Store Settings...' is displayed briefly.
'-> Settings' is then displayed.
[STOP/OPEN] button once to exit the '-> Settings' menu
10.
Press the
or
Press the
[STOP/OPEN] button twice to exit the '*MACHINE MENU*' .
The indicator backlight flashes as a visual signal after the centrifugation run
is finished.
1.
Press and hold the
'*MACHINE MENU*' is displayed after 8 seconds.
��
[SELECT] button repeatedly until '-> Settings' is displayed.
2.
Press the
3.
Press the
[START/PULSE] button.
'End beep = on' or 'End beep = off' is displayed.
��
Rev.: 10 / 11.2023
[SELECT] button.
[SELECT] button.
Software operation
29 / 39

Werbung

loading

Diese Anleitung auch für:

18011801-01