mode (refer to the measurement mode setting) and use specified cuff for pediatric or neonate.
Using incorrect measurement mode may cause danger to the patient, because the adult
pressure level is relative high and is not suitable for pediatric or neonatal patients.
In neonatal mode, the initial inflation value of cuff is 9.3kPa (70mmHg). The cuff can be
inflated to 20kPa (150mmHg), if the inflating value exceeds this value, the device will reset
and deflate automatically.
The device has double overpressure protection for hardware and software. If over-inflating
occurs, the device will reset and deflate immediately. If the device keeps over-inflating status,
please disconnect the cuff from the device, and cut off the power or turn off the device.
Please use the device under proper temperature and humidity (refer to Specification),
otherwise measured result may not accurate.
Note
Measure should be taken in a quiet place,and relax the body.
Remain still 4~5 minutes before measure.
Relax the body, do not let the muscle function.
Do not talk and move during the measure.
Wait 4~5 minutes when measuring in succession.
Do not use mobile equipment such as mobile telephone near the device.
37