Operation
This warning message can be acknowledged at any time, and this has no negative effects on
the ongoing processes of the magnetic stirrer.
Limiting the travel distance
The travel distance of the lab jack can be limited mechanically with the help of suitable and
adjustable clamping levers (optional accessory):
Each adjustable clamping lever [1] is mounted with a suitable shim [2] and a guide piece [3]:
Place the guide piece [3] in the guide of the lab jack [4a] from the inside.
Place the shim [2] on the thread of the clamping lever [1].
Screw the clamping lever [1] onto the guide piece [3]:
Fix the second clamping lever in the same way in the opposite guide of the lab jack
[4b].
The position of the clamping lever can be adjusted, if necessary:
Pull the clamping lever outwards to unlock it [1a].
Turn the clamping lever clockwise or counterclockwise in a suitable position [1b].
Move the lab jack upward to the highest working point [5], push the two clamping levers
onto the two guide pins [6] until they stop and tighten the two clamping levers.