Herunterladen Diese Seite drucken

OHAUS 3000-Serie Bedienungsanleitung Seite 164

Indikatoren
Vorschau ausblenden Andere Handbücher für 3000-Serie:

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

EN-18
3.4
Setup Menu
When the Indicator is used for the first time, enter this menu to set the
Capacity and Graduation.
3.4.1 Reset
Reset the Setup menu to the factory defaults.
No
= not reset.
Yes
= reset.
3.4.2 Legal for Trade
Set the legal for trade status.
OFF
= off
ON
= on
Turning on the "LFT" menu setting has the following effects:
• Zero-range is set and locked on "2".
• Auto Zero Tracking is set and locked on 0.5d
• The lb:oz unit is not available as a power-on setting.
3.4.3 Calibration Unit
Set the unit during calibration.
CAL UN kg = Calibrate using kg weights
CAL UN lb = Calibrate using pound weights
3.4.4 Capacity
Set the scale capacity from 5 to 20000. Refer to the Setup Table 3.3 for available settings.
3000 Series Indicators
Reset
No, Yes
Off, On
Legal for Trade
kg, lb
Calibration Unit
5...20000
Capacity
Graduation
0.0005...20
Power On Unit
Auto, kg, lb, g, oz
2%, 100%
Zero Range
Off, On
Retain Zero Data
End Setup
Exit SETUP menu

Werbung

loading

Diese Anleitung auch für:

T32meT32mc