Herunterladen Inhalt Inhalt Diese Seite drucken

Temperature Tables - Endress+Hauser promass 64 II2G Ex-Dokumentation Zu Den Betriebsanleitungen

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

en
Description of Promass I
DN 16 = DN 15 "FB"
DN 26 = DN 25 "FB"
DN 41 = DN 40 "FB"
"FB" = Full bore
Note!
4

Temperature tables

Compact version (Sensor)
at T
= 50 °C
a
Promass F DN 8/15/25/40/50/80/100
Promass M DN 8/15
Promass M DN 25/40/50
Promass M DN 80
at T
= 60 °C
a
Promass A DN 2/4
Promass F DN 8/15/25/40/50
Promass F DN 80/100
Promass M DN 8/15
Promass M DN 25/40/50
Promass M DN 80
The minimum fluid temperature is -50 °C (for a minimum fluid temperature below -20 °C,
the minimum ambient temperature is -10 °C)
Remote version (Sensor)
at T
= 50 °C
a
Promass M DN 8/15
at T
= 60 °C
a
Promass A DN 2/4
Promass F DN 8/15/25/40/50
Promass F DN 80/100
Promass M DN 8/15
Promass M DN 25/40/50
Promass M DN 80
Promass A DN 2/4
Promass F DN 8/15/25/40/50
Promass F DN 80/100
Promass M DN 8/15/25/40/50
Promass M DN 80
Note!
With the fluid temperatures given, and for a certain temperature class, the compo-
nents will not be subjected to any non-permissible temperatures.
Transmitter
In the PRO-LINE EEx d housing, the PROMASS 64****-*****B***** resp.
PROMASS 64****-*****C***** transmitter has the T6 temperature class up to an
ambient temperature of T
Max. fluid temperature [°C] in
T6
T5
60
95
60
95
65
95
65
80
Max. fluid temperature [°C] in
T6
T5
60
95
60
95
60
75
95
65
95
65
80
Max. fluid temperature [°C] in
T6
T5
55
95
Max. fluid temperature [°C] in
T6
T5
60
95
60
95
60
75
95
65
95
65
80
Min. fluid temperature [°C] in
Use in gas group IIC
–90
–120
*
–50
*
= 60 °C.
a
Promass 64
T4
T3
T2
130
170
200
130
150
130
150
110
150
T4
T3
T2
130
150
200
100
100
100
100
100
T4
T3
T2
130
150
T4
T3
T2
130
150
200
130
170
200
110
170
200
130
150
130
150
110
150
Use in gas group IIB
–120
–120
–50
–50
–50
* not available for this gas group
Endress+Hauser

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis