Promass 64
Sensor A
Sensor F
Sensor M
Transmitter Promass 64 in:
EEx de [ib] IIB/IIC or EEx d [ib] IIB/IIC resp.
EEx de [ib/ia] IIB/IIC or EEx d [ib/ia] IIB/IIC
PRO-LINE Ex d housing in:
EEx d resp. EEx de
(an appropriate cable entry is to be selected,
depending on the type of protection)
Sensor in EEx ib IIC:
PROMASS A
DN 2...4
PROMASS F
DN 8...50
PROMASS M
DN 8...50
The maximum permitted cable length is 120 m
when using cable type (6 Li9YCY 0.38 mm
however for signal reasons, the maximum cable length
is limited to 20 m.
For ambient and fluid temperature ranges,
and temperature class, see page 4
Endress+Hauser
Hazardous area
II2G (ATEX)
Zone 1 (CENELEC)
Hazardous area
2
) FCY,
II3G (ATEX)
PROMASS 64
PROMASS 64
PROMASS 64
Zone 2 (CENELEC)
Transmitter Promass 64 in:
EEx de [ib] IIB/IIC or EEx d [ib] IIB/IIC resp.
EEx de [ib/ia] IIB/IIC or EEx d [ib/ia] IIB/IIC
PRO-LINE Ex d housing in:
EEx d resp. EEx de
(an appropriate cable entry is to be selected,
depending on the type of protection)
Sensor in EEx ib IIB:
or
PROMASS F
DN 80...100
PROMASS M
DN 80
The maximum permitted cable length is 120 m
when using cable type (6 Li9YCY 0.38 mm
however for signal reasons, the maximum cable length
is limited to 20 m.
For ambient and fluid temperature ranges,
and temperature class, see page 4
en
Safe area
Safe area
2
) FCY,