Basic safety instructions
6.6
Product safety
6.6.1
State of the art
The product is designed to meet state-of-the-art safety requirements, has been tested, and
left the factory in a condition in which it is safe to operate. The relevant regulations and
European standards have been observed.
6.6.2
Electrical equipment in hazardous areas
When used in conjunction with the CYK10-E/G**1/2 or CYK20-BA**** measuring cables,
the Memocheck CYP02D-***G sensor simulators, which are certified in accordance with
ATEX, may be connected only to the certified, intrinsically safe, digital sensor output circuits
of the following field measuring devices:
Liquiline M, type CM42-ME******** or CM42-MF******** in accordance with EU type-
examination certificate BVS 04 ATEX E 121 X.
Memosensis approved for measuring applications in explosive atmospheres in accordance
with:
• IECEx design approval, IECEx BVS 12.0007
• ATEX design approval BVS 12 ATEX E 008 with amendments
The EU Declaration of Conformity is an integral part of these Operating Instructions.
When used in conjunction with the CYK10-G**1 measuring cable, the Memocheck CYP02D-
***G sensors simulators, which are certified in accordance with FM / CSA, may be connected
only to the certified, intrinsically safe, digital sensor output circuits of the following field
measuring devices:
Liquiline M, type CM42-MP******** or CM42-MS******** as per the control drawings for
the transmitter.
Operating Instructions Liquiline M CM42, Part 2, BA00382C
The electrical connection of the CYK10 or CYK20 measuring cable must be carried out in
accordance with the wiring diagram for the transmitter.
The sensor simulators must not be operated under electrostatically critcical process
conditions. Avoid strong steam or dust currents that act directly on the connection system.
The maximum permissible cable length is 100 m (330 ft).
Hazardous area versions of digital sensors and simulators incorporating
Memosenstechnology are indicated by a red-orange ring on the plug-in head.
Temperature classes
Simulator
Memocheck
CYP02D-***G
If the ambient temperatures shown above are not exceeded no invalid temperatures for the
particular temperature class will occur at the simulator.
12
Ambient temperature range T
-15 to +70 ˚C
Memocheck CYP02D
Temperature class
a
T6