Vibration velocity
mm/s
> 11.0
> 0.43
7.1 ... 11.0
0.28 ... 0.43
4.5 ... 7.1
0.18 ... 0.28
3.5 ... 4.5
0.14 ... 0.18
2.8 ... 3.5
0.11 ... 0.14
2.3 ... 2.8
0.09 ... 0.11
1.4 ... 2.3
0.06 ... 0.09
< 1.4
< 0.06
Vibration severity zones for vibration velocity according to DIN ISO 10816-3
Vibration severity zones for vibration displacement according to DIN ISO 10816-3
5.1.4
Display max value
This sub menu is used to activate the display of the max value. When enabled, the highest reading
so far is displayed separately below the current measurement value. The BACK
used to reset the max value.
5.2
Data logger
In this menu, the data logger can be started and the data logger settings can be changed.
5.2.1
Start data logger
The data logger is started via this sub menu which opens the data logger screen where the current
measuring parameters, the reading and the data logger settings are displayed. For the data
logger, the same settings as for the general measuring mode are used. These can be changed
as described under 5.1 Measurement.
Group 1
in/s
rigid
D
C
B
A
© PCE Instruments
flexible
D
C
B
A
42
Group 2
rigid
flexible
D
D
C
C
B
B
A
A