6.8 Linearization (non-verified models only)

Linearity shows the greatest deviation of a weight display on the scale to the value of
the respective test weight according to plus and minus over the entire weighing
range.
If linearity deviation is discovered during a testing instrument control, you can
improve this by means of linearization.
Carrying out linearization is restricted to specialist staff possessing well
acquainted with the workings of weighing scales.
The test weights to be used must be adapted to the weighing scale's
specifications; see chapter 3.4 "testing instruments control".
Observe stable environmental conditions. Stabilisation requires a certain
warm-up time.
After successful linearization you will have to carry out calibration; see
chapter 3.4 "Testing instruments control"
Tab. 1: Adjustment points
Adjustment
weight
1.
2.
3.
4.
Operation
How to carry out linearization:
 Switch on balance and during self-check
press the TARE key
 Use the numeric keys to enter password
„9999"
 Confirm with TARE button
20
CPB 6K0.1N
CPB 15K0.2N
1 kg
2 kg
4 kg
6 kg
CPB 30K0.5N
2.5 kg
5 kg
10 kg
15 kg
Display
CPB-N / CPB-DM-BA-e-1831
5 kg
10 kg
15 kg
30 kg