p p p p p Names of parts
Anti-dust bellows
Feeler (Interchargeable)
Stem (Diameter: 8 mm/0.31")
• Illustration: DG810BL
• The names of the parts for DG805BL are the same as those for
DG810BL.
• The air lifter cannot be attached to the DG810B.
p p p p p Specifications
Item
Measuring range
Resolution
1)
2)
Accuracy
Maximum response speed
Downward
Measuring
forces
2)
Horizontal
Upward
Operating temperature
Storage temperature
Protection against dust
Cable length
Stem diameter and length
Mass
3)
Measuring feeler
Life time
Note
1) As connected to LT20, LY51/LY52, LG10, LH51/LH52 or LH61.
2) Accuracy and measuring forces given above are those obtained at 20 °C/ 68 °F.
3) The mass indicated is the total mass excluding the cable and interpolation box.
2 (E)
Model
DG805BL
0 to 5.2 mm (0 to 0.20")
0.45±0.25 N
0.4±0.25 N
0.35±0.25 N
0 °C to 50 °C/ 32 °F to 122 °F (Interpolation box: 0 °C to 40 °C/ 32 °F to 104 °F)
φ8
mm ×74.5 mm
+0
–0.009
φ0.31
" mm ×2.93" mm
+0
–0.0004
Approx. 20 g/0.71 oz
DZ-123 provided with carbide ball tip (Mountscrew M2.5)
Air Lifter Connector
Interpolation box
Connector
DG810B
0.0005 mm
0.002 mm (0.00008")
1 m/s
–20 °C to 60 °C/ –4 °F to 140 °F
by bellows
5 m/ 16'
φ8
mm ×114 mm
+0
–0.009
φ0.31
" mm ×4.49" mm
+0
–0.0004
Minimum 5 million cycles without shock
Cable (5m/16')
DG810BL
0 to 10 mm (0 to 0.39")
0.6±0.3 N
0.5±0.3 N
0.4±0.3 N
φ8
mm ×112 mm
+0
–0.009
φ0.31
" mm ×4.41" mm
+0
–0.0004
Approx. 40 g/1.41 oz
DG810B

Werbung