Display Indications
Display
BAt
FLo
Out
Outc
Err2
500
ACE_A_BA_20190221.indd 22
Meaning
The battery is empty.
The testing cannot
be performed
because the battery
is low.
The sampling
volume or blow
power is not
enough to analyze
the breath sample.
If no breath sample
was given within
3 minutes, the
device turns off
automatically.
The testing cannot
be performed
because the device
is turned on outside
of the operating
temperature.
"Err2" states that
fluid or saliva has
entered the device
and the sensor is
unable to evaluate it.
The total number
of breath tests per-
formed has reached
500. The device
should be calibrated
promptly to maintain
the warranty and
accuracy.
Action
Replace with new
batteries.
Replace with new
batteries.
Press the start button, the
countdown will start again,
and "BLOW" will appear
on the display. You can
now perform a new test.
Press power button to
restart the countdown
cycle and retest.
Repeat the test after
the unit has reached
a temperature within
the operation range
(5°C to 40°C).
Wait 20 minutes and
then try again. If "Err2" is
displayed for a long time,
contact your seller for
customer service.
Contact your seller for a
calibration service. You
can send the device for
calibration to the service
address specified at
www.alkomat.net.
22